CARDIAC STIMULATOR
ENGLISH Page 35 of 49
12.8.12 Specialised Protocols
The following specialised protocols are available from the Procedure Menu (‘m’). They are self-
explanatory.
Bi Ventric Pacing. This protocol requires the Four Channel Upgrade, using the Four Channel
Multiplexer Kit, MP3091:
Bi Ventricular
('m,e')
S 1
S1
S1
S2
S3
Ch1-Atr:
Ch2-L V:
Ch3-RV:
S2
S3
S2
S 3
S3= 50 S3= 0 S3= -50
S1
S2
S3
S1
S 2
S3
Paced S2. S2 into a different channel to S1 is interpolated between S1stimuli. S2 may be 0 or –ve,
meaning it coincides or preceedes the last S1:
S2
Paced_S2
('m,i')
S1 S 1 S1 S1
S1 S1
Train = 4
S2
S1 S1 S1
S2
S1
S2=5 0
S2=20 S2= -2 0
Sensed S2&S3. Multiple sensed extra-stimuli are given into different channels / chambers:
S2 & S3 i nto different cham bers
(s ens ed )
TRAIN = 3
S2
P AUSE
Sensed_S2&S3
('m,j')
t t t t
t
S3
(se ns ed)
TRAIN = 3
S2
t t t
S 3
Ch1-Atr:
Ch2-Vent:
S3 =
-
250 to +250m s
S3 = 60 S3 = 30