CARDIAC STIMULATOR
ENGLISH Page 41 of 49
Software Related
Advisory
Messages
Explanation Suggested Action
xxx % RR = xxx In Burst and Overdrive protocols indicates
calculation of S1 as a percentage of
measured RR intervals.
Use calculated interval if correct.
Lower Limit xxx Attempt to set S1, Decrement or Train
parameter BELOW limit set in
Configuration Menu.
Enter new value within limits.
Upper Limit xxx Attempt to set S1, Decrement or Train
parameter ABOVE limit set in
Configuration Menu.
Enter new value within limits.
Timing Error Inaccuracy in verification between PC and
SGU clocks.
Stimulus timing may be incorrect
– verify stimulation timing with
third party equipment. Service the
Micropace Stimulator system.
Unstable RR RR interval not stable during adaptive
calculation of S1 interval in ATP protocol.
Check & improve ECG signal /
QRS detection.
Output
Interference
Outputted current out of tolerance of 25%
for either channel; may be due to RF
Energy interference on output.
Remove interference; if persists,
service SGU.
Table 7 Advisory Messages and suggested actions.
12.13.2 Text Messages
Advisory Notes
(NOTE…)
These messages inform user of required instructions or informative
data and are self-explanatory
They represent either normal or correctable abnormal run-time
program conditions.
They require a user response and do not terminate the program.
Run Time Warnings
(WARNING…)
These messages warn user of possibly incorrect or inappropriate
actions by him.
They require user action or confirmation of an action and do not
terminate the program.
Data Error Messages
(DATA_ERR…)
File Error Message
(FILE_ERR…)
Run Time Error
Messages
(RUN_ERR…)
Significant errors found, stop using the Stimulator and contact your
distributor or Micropace for service.

Table of Contents