12.12 Software Error Messages
Error messages have help messages associated with them. In some cases help text appears with the
error message; with other errors, pressing 'h' displays the associated help text.
Run time timing error
The following message may appear while using the Stimulator.
Internal Error: STIM_OVERRUN: The Computer is not keeping up, stimulus timing may be inaccurate!
This can happen if you have continuously pressed a key, forcing the software to service the keyboard
too frequently, or there is some problem with the computer. Note that subsequent pacing will most
likely be accurate, but this message will come up only once in any one session and any repetition of
this error will be ignored for the rest of the session, to enable you to continue pacing the patient if
necessary. To re-enable this error detection, quit the program with 'Q' and follow the prompts to restart
the software or by rebooting the computer.
Program warnings on exit
The following message may be displayed after exiting the program.
Program exited with Warnings, logged in file 'stim.log'.
This indicates that a warning was issued during execution of the program.
Contact your Distributor for further details if required. Abnormal program exit will also result in the
display of a Recovery Menu. Follow the suggested menu items to recover your last valid configuration
settings or to re-install the program.