USER INSTRUCTION MANUAL
Page 24 of 49 ENGLISH
10.4 Setting up the computer
a. Switch on (i) Isolation transformer, (ii) Bona PC (push front button once), (iii) LCD Display (on
the side) and (iv) SGU at the rear. Allow system to boot up.
b. When prompted respond that you are the distributor (to avoid having the License agreement
shown to you)
c. If you have an EU version, you will be offered a menu to chose an interface language.
d. When prompted to calibrate screen, touch screen in the places indicated by crosses from the
seated position and the same eye level as the customer will use, using the stylus. Calibration
of each touch screen must be performed.
e. You will now see the Main Stimulator Screen.
f. The Stimulator software will already be configured for your hardware setup – touch screen,
four channels and one external ECG
10.5 Indicating location of SGU
Where the SGU, MP3008, is located out of sight of the operator’s normal position, apply label MC0809
“Backup Stimulator Here” to make SGU’s location clearly visible to the operator during normal use.