12.8 Pacing Protocols
The Micropace Stimulator features a number of pre-programmed stimulation protocols.
12.8.1 Protocol Selection & Renaming
Favourite protocols in the Protocol Toolbox (‘M’ hot key) may be selected with checkbox to appear
in main Protocol Menu and may be renamed from a customizable list using the EDIT key.
Figure 14 Protocol Toolbox
12.8.2 Pace Protocols
S1
PACE
('p')
Select the PACE protocol to pace at a fixed interval set by S1 into selected pacing site.
12.8.3 Threshold Protocol
THRESHOLD
('t')
S1
Reducing current
Set Current = 2 * threshold, (>1.0mA)
TRAIN=4
User
pacing
stops
The THRESHOLD protocol helps to determine the pacing threshold by continuously decreasing or
increasing the pacing Current amplitude; the Operator needs to stop pacing when capture is lost, and
accept or modify a displayed automatically calculated new default current (twice threshold current, and
> 1.0 mA).