CARDIAC STIMULATOR
ENGLISH Page 47 of 49
Error
Message
Error
Name
E
xplanation
Suggested action
ablation voltage). This may cause
unwanted arrhythmias.
Stimulus Generator Unit by pressing
'Backup Enable'. If error persists
contact distributor.
_,t,o,L ,E,r,r
Tolerance
Error
Pulse Tolerance Error. More that
200 pulses were out of the +/- 25%
tolerance limit.
Remove any interference from the
Stimulator outputs. Reset Stimulus
Generator Unit by pressing 'Backup
Enable'. If error persists contact
distributor.
N,o,P,C,E,r,r
No PC
Error
No communication received from
PC for a while. Due to failure of PC
or disconnection of the
communication cable.
Check that communication cable
between PC and Stimulus Generator
Unit is secure. Restart both the
Stimulus Generator Unit and the PC
by switching them off and on.
r,E,r,r , E,r,r
Receive
Error
Serial RS232 data link receive
errors
Noise on the data link between
Stimulus Generator Unit and the PC.
Check the cable.
Table 9 Tabulated Error messages