EasyManua.ls Logo

Miele PLW 8616 - Program Overview

Miele PLW 8616
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Application technology
34
Program overview
At delivery, the reprocessing programs are assigned to the A 503mobile unit. The follow-
ing modules can be used in various combinations:
- A 612
- A 620
- A 621
- A 622
Exceptions to this are the Vials and Pipettesprograms (see table).
In order to use other mobile units, new load carriers must be set up in the web interface
under “Load carrier management”. “Load carrier management” is described in the pro-
gramming manual.
Power levels
The mobile units in the modular loading concept can be fitted with modules in different
ways. Depending on the type and number of modules used, different volumes of water and
pump speeds are required to achieve a similar wash pressure. The individual reprocessing
programs are therefore available with different power levels. The higher the power level of a
reprocessing program, the higher the volume of water and the pump speed.
The program structure and the following parameters, which are relevant to performance,
are identical in the different power levels:
- Water quality
- Temperature
- Holding time
- Dispensed process chemical in %
- Drying temperature
- Drying duration
(see appendix: “Power levels/mobile units/module combinations” table)