READY Mode
– water quality
values
• The READY Mode screen display is explained below.
• This screen shows the resistivity and temperature of the water filling the
Reservoir.
• The Reservoir water is further purified and dispensed from the Q-POD Unit.
• The Q-POD Display shows the quality of this water.
READY Mode screen Explanation
In this example, the water filling the
Reservoir has:
• a resistivity of 12.5 MΩ.cm,
• is temperature compensated,
• a temperature of 25.1°C, and
• the Reservoir is 80% full.
NOTE:
When the Milli-Q System stops
filling the Reservoir, the last
measurements of water quality are
displayed.
• In this example, the Milli-Q
System was powered on but did
not begin to fill the Reservoir.
• In this case, there are no water
quality measurements to display.