Summary of Alarm messages
Alarm
messages
LCD message What it means
A10 ERROR 0
A10 PCB E
2
Prom defective.
Power off, power on the Milli-Q
System.
Dispense water for several
minutes.
If the message continues, then
contact Millipore.
A10 ERROR 1
A10 PCB A/D converter
defective.
Power off, power on the Milli-Q
System.
Dispense water for several
minutes.
If the message continues, then
contact Millipore.
A10 ERROR 2
A10 Thermistor defective.
Power off, power on the Milli-Q
System.
Dispense water for several
minutes.
If the message continues, then
contact Millipore.
A10 ERROR 3
Problem occurred with
temperature compensation.
Power off, power on the Milli-Q
System.
Dispense water for several
minutes.
If the message continues, then
contact Millipore.
A10 ERROR 4
The water entering the A10 is <
4°C.
If the problem can not be resolved,
then contact Millipore.
A10 ERROR 5
The water entering the A10 is >
41°C.
If the problem can not be resolved,
then contact Millipore.