Summary of Alarm messages, Continued
Alarm
messages
(continued)
LCD message What it means
A10 ERROR 6
The conductivity of the water
entering the A10 is > 1.1 μS/cm.
If the problem can not be resolved,
then contact Millipore.
A10 ERROR 7
The temperature inside the A10
during its Analysis Mode
exceeded 55°C.
Power off, power on the Milli-Q
System.
Dispense water for several
minutes.
If the message continues, then
contact Millipore.
A10 ERROR 8
The TOC sample oxidation was
not completed in the allotted time.
If the message continues, then
contact Millipore.
A10 ERROR 9
The A10 is not detecting a TOC
value.
This can be caused by:
The A10 Solenoid Valve is not
closing and could have a particle
stuck in it or
The A10 Lamp is not turning on.
Perform an A10 Cleaning Mode.
This might dislodge a stuck
particle
Replace the A10 Lamp if the A10
Cleaning did not fix the problem.
If the message continues, then
contact Millipore.
CHECK A10 COM
The communication between the
A10 TOC Monitor and the Milli-Q
System PC Board is interrupted.
The TOC value is no longer
reported.
Contact Millipore.