Calibrating the Flowrate
Introduction
The Milli-Q Water flowrate should be calibrated when the Milli-Q System
is installed.
This calibration should be done with Q-POD Unit 1 in case there are
multiple Q-POD Units.
A 1 Litre graduated cylinder is needed.
Procedure
Follow the steps below to perform a Flow Calibration.
Step Action Diagram
1 Go to STANDBY Mode.
2
Select <Menu>.
Press .
3
Enter the <Manager Menu>.
See the Software Chapter to
learn how to enter the
Manager Menu.
4
Select <Setup>.
Press
.
Continued on next page