Electricity and fire hazards
Potential risks arising from electrical devices include electric shock, short, and overheating. Electric shock may
lead to severe personal injury or death. An electric short and overheating may lead to the spread of fire.
Electronic equipment might emit sparks, which could ignite combustible vapors or material resulting in
explosion or fire.
Do not use the instrument in areas classified as hazardous locations; for example, in oxygen-
laden environments. If flames or smoke appear, immediately switch off the power supply,
unplug the instrument from the electrical outlet, and contact an authorized Miltenyi Biotec
service provider or Miltenyi Biotec Technical Support.
The instrument housing is designed to lower the risk of electric shock, short, and spread of fire.
Except for the front access covers, do not remove or penetrate any cover, nor any cover of
other accessory hardware supplied by Miltenyi Biotec. Only authorized personnel is allowed to
remove any other cover of the instrument. Never push an foreign object through an opening
into the instrument.
Do not use the instrument if
l it has been opened or disassembled
l it has been dropped or damaged
l it has damaged or broken parts
l it has a damaged power cable
l an object has been entered through the ventilation slots
l an object has been dropped into the instrument
The instrument is intended for indoor use only. If liquids enter the instrument, it may lead to electric shorts,
shock, or fire. Take special care while handling liquids.
Do not allow liquids to enter the interior of the instrument. Protect the instrument against
accidental spillages and splashes. Clean up spillages immediately. Do not operate the
instrument if liquids have entered the instrument. Do not use the instrument in a wet and damp
location. Avoid areas with high humidity or condensation. Moving the instrument from a cold environment
such as a cold room at 5°C to room temperature may cause condensation inside the instrument. In such
cases, wait for the instrument to dehumidify before operating it. Before cleaning, unplug the instrument. Do
not use liquid or aerosol cleaning agents, always use a damp cloth.
Protect the instrument from overheating. Do not use the instrument in areas with more than 25°C. Ensure
adequate air circulation in the room when the instrument is operated. Allow sufficient air circulation around the
instrument at least 15 cm on all sides during operation. Do not place the instrument near radiators,
heat registers, stoves, or other equipment (including amplifiers) that generate heat. Protect the
instrument from direct sunlight. Do not cover the slots and openings of the instrument as
these are provided for ventilation. Do not place the instrument in a built-in rack or similar confined
spaces unless the space has been specifically designed to provide proper ventilation.