Accutorr Plus™ Operating Instructions 0070-10-0692-02 6 - 11
Appendix Electromagnetic Compatibility
TABLE 6-2
GUIDANCE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The Accutorr Plus is intended for use in the electromagnetic environment specified below. The
customer or the user of the Accutorr Plus should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines ±1
kV for input/
output lines
±2 kV for power
supply lines ±1
kV for input/
output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV
differential
mode ±2 kV
common mode
±1 kV differential
mode ±2 kV
common mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
<5% U
Τ
(>95%
dip in U
T
) for
0.5 cycle
40% U
T
(60%
dip in U
T
) for
5 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
<5% UT (>95%
dip in U
T
) for 5
sec.
<5% U
Τ
(>95%
dip in U
T
) for 0.5
cycle
40% U
T
(60%
dip in U
T
) for
5 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
<5% UT (>95%
dip in U
T
) for 5
sec.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Accutorr
Plus requires continued operation
during power mains interruptions, it is
recommended that the Accutorr Plus be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
U
T
is the A.C. mains voltage prior to application of the test level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals