Accutorr Plus™ Operating Instructions 0070-10-0692-02 4 - 5
User Maintenance Battery Maintenance and Replacement
4.4.2 Battery Replacement
1. Open battery compartment door, located on the bottom of the unit, by pressing the
release lever.
2. Press the release lever, located next to the battery, to eject the battery. Slide out the old
battery.
3. Slide in the replacement battery until it clicks into place.
4. Close the battery compartment door.
5. Batteries are shipped partially charged and therefore require charging prior to use. It is
recommended that the Lithium ion battery be charged in a Accutorr Plus for 4 hours
minimum prior to use.
CAUTION: Removing the battery from the Accutorr Plus while the AC
line cord is disconnected may cause the alarm settings to be
reset to their defaults.