Specifications Appendix
6 - 8 0070-10-0692-02 Accutorr Plus™ Operating Instructions
6.2.13 Electrical Ratings
6.2.14 Safety Characteristics
Risk (Leakage) Currents
Dielectric Withstand
1500 V RMS at 50 or 60 Hz for 1 minute from AC mains hot or neutral to chassis
2500 V RMS at 50 or 60 Hz for 1 minute from any patient lead or combination of patient
leads to chassis
NOTE: These two tests satisfy IEC 60601-1 requirements for double
or reinforced insulation (tested at 4000 V RMS) between the
applied part and live parts (between patient leads and AC
mains hot or neutral)
Ground Resistance
0.1 ohm from the AC mains power inlet module’s ground contact pin to any exposed
metal part which may become energized when measured per IEC 60601-1. A ground
resistance of =0.2 ohm is allowed when measured from the U blade of the supplied AC
line cord to any exposed metal part which may become energized.
Voltage: 100 - 120 / 220 - 240 VAC
Current: 0.85 / 0.5 A
Frequency: 60 / 50 Hz
Power Consumption: 40 W, maximum
Enclosure Risk (Leakage)
Current: 100µA normal operating conditions
300µA single fault condition
Patient Source Current: 10µA normal operating conditions
50µA single fault condition
Patient Sink Current: 50µA.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals