Accutorr Plus™ Operating Instructions 0070-10-0692-02 6 - 3
Appendix Specifications
Inflation Source
This inflation source is capable of supplying sufficient air to bring a volume of 700cc’s to a
pressure of 300 mmHg in no more than 35 seconds. If the cuff is not inflated to the desired
pressure within 60 seconds then the cuff is vented and a retry cycle is initiated.
Leak Rate
With the bleed valve closed, the maximum pressure drop shall be 10 mmHg in 90 seconds
measured with a 700cc volume at a differential pressure of 250 mmHg.
Cuff Vent Rate
When the unit is vented, a volume of at least 700 cc’s is reduced from a pressure of 250
mmHg to a pressure of 20 mmHg in a maximum of 14 seconds.
6.2.6 Temperature (Predictive)
6.2.7 SpO
2
- Nellcor
®
Performance Specifications
Range: 90-110°F, 32-43°C
Display Resolution: 0.1°F, 0.1°C
Accuracy: Meets ASTM E1112-86 for accuracy
Range: 70-100% SpO
2
Display Resolution: 1% SpO
2
SpO
2
Response Time: 4.5 to 6.5 seconds
Display Update: Less than 4 seconds
Calibration: Factory Calibrated to Functional Saturation
Accuracy - Nellcor
®
:* ± 2 digits from 70 - 100% SpO
2
- Adult
±3 digits from 70 -100% SpO
2
- Neonates
<70% unspecified
Pulse Rate Range - Nellcor 21 - 249 BPM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals