Battery Maintenance and Replacement User Maintenance
4 - 4 0070-10-0692-02 Accutorr Plus™ Operating Instructions
4.4 Battery Maintenance and Replacement
4.4.1 Battery Maintenance
The Accutorr Plus is available with a lithium ion battery. This battery type may be subject to
local regulations regarding disposal. At the end of the battery life, dispose of the battery in
accordance with any local regulations.
CAUTION: Batteries used in this device may present a risk of fire or
chemical burn if mistreated. Do not disassemble, heat above
100°C (212°F), or incinerate. Replace Lithium Ion battery
with P/N: 0146-00-0069 only. Use of another battery may
present a risk of fire or explosion.
CAUTION: Dispose of used battery promptly in accordance with local
laws. Keep away from children. Do not disassemble and do
not dispose of in fire.
CAUTION: Recharge the Lithium ion battery while in the unit at room
temperature. If the Accutorr Plus is being used in a hot
environment, the Lithium Ion battery may not charge when
the unit is connected to AC. This safety feature is important
because charging a hot battery shortens the battery’s life
span.
CAUTION: Remove the battery if the Accutorr Plus is not likely to be
used for an extended period of time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals