To View and Delete Stored Data (Trend Mode) Operation
3 - 14 0070-10-0692-02 Accutorr Plus™ Operating Instructions
3.6 To View and Delete Stored Data (Trend Mode)
The Accutorr Plus is capable of storing up to 100 entries of measurement data. Each time a
successful NIBP measurement is made, the data is automatically stored in memory. When a
temperature measurement is made between two minutes before and two minutes after an
NIBP measurement, it is stored as the same entry with the NIBP measurement. If a
temperature measurement is made outside this time, it is stored as a separate entry. When
either NIBP or temperature measurements are stored and SpO
2
information is available, then
the SpO
2
data is also stored.
When 80 to 99 entries are stored into trend memory, the MEMORY FULL Indicator (20) will
flash. When 100 entries are stored into trend memory, the MEMORY FULL Indicator (20) will
illuminate continuously. Once 100 entries are stored, old data can be deleted manually for
any patient; or when new data is available, the Accutorr Plus will automatically delete the
oldest data for the currently displayed patient.
NOTE: The unit will also automatically delete data that is 24 hours
old.
The Accutorr Plus basic model uses the Systolic, Diastolic, MAP, and Temp displays to view
stored data. The Accutorr Plus advanced models display up to 5 measurements at a time. The
stored data that is viewed is for the currently selected patient (indicated by the room number/
bed letter).
3.6.1 To View the Stored Measurements on the Accutorr Plus, basic model
1. Press the DATA SCAN key (22) (1 beep tone). The DATA SCAN Indicator (23)
illuminates.
2. Press the PATIENT INFO. Up and Down Arrow keys (27 & 28) to view stored data
for the current patient. The stored data is displayed in the Systolic, Diastolic, MAP, Pulse
Rate and Temp displays.
Consecutive presses or pressing and holding the UP or DOWN arrow will allow the
stored measurements to continuously wrap around. When the measurements wrap, a
double beep tone will sound. If a temperature measurement is not available for the NIBP
measurement that is displayed, then - - - - is shown in the Interval/Elap. Time/Temp
display (33). To view the time of measurements, press the TIME/TEMP key (32).
3. To exit the view stored data mode, press the DATA SCAN key (22) (1 beep).
3.6.2 To View the Stored Measurements on the Accutorr Plus, advanced
models
The stored measurements on the Accutorr Plus advanced models are displayed in the LCD.
Up to 5 stored measurements are displayed at one time. Measurements are displayed in time
order, with the newest measurement at the top. A scroll bar with one or both arrows will
display on the right side of the LCD when more measurements are available to view. When
only one arrow displays, more measurements are only available in the direction of the arrow.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals