Status and Error Codes Operation
3 - 36 0070-10-0692-02 Accutorr Plus™ Operating Instructions
3.16 Status and Error Codes
The Accutorr Plus uses the various displays on the front panel to display the operational
status. Status and error codes listed below can generally be resolved by the user however,
some error codes, which are marked with an asterisk (*), may require resolution by a
qualified technical service person. These codes with their descriptions are listed on the back
of the Quick Reference card.
NOTE: Status codes 8810 through 8858 can be cleared from the
Room and Bed displays by pressing the Room/Bed key (24).
Status and Error Code Table
TYPE CODE DESCRIPTION REASON
NIBP 8810 Retry - Unable to Measure Motion artifact, cycle time-out, weak
pulsations or no pulsations. A triple beep
tone is generated.
8811 Retry - Pump Higher Insufficient cuff pressure. A triple beep tone
is generated.
8812 Stop - Cuff Overpressure Excessive cuff pressure detected by the
software. A triple beep tone is generated.
8813 Stop - Unable to Measure 4 successive measurement attempts failed.
A triple beep tone is generated.
TEMP
(PTM)
8830 Check Probe Tissue contact may have been lost.
8831 Replace Probe Defective probe or connection.
SpO
2
8850 No Sensor No sensor connected.
8851 Sensor Off Sensor not on patient. (Masimo SpO
2
only)
8852 Interference Interference on signal. (Masimo SpO
2
only)
8853 Pulse Search Unit cannot find signal. (Nellcor SpO
2
Module will report “Pulse Search” -8853-
when the sensor is not on the patient.)
8854 Weak Pulse Weak pulse detected. (Masimo SpO
2
only)
8856 Check Sensor Sensor problem. (Masimo SpO
2
only)
8857 PR<21 Pulse rate is less than 21 bpm. (Nellcor
SpO
2
only)
8857 PR<26 Pulse rate is less than 26 bpm. (Masimo
SpO
2
only)
8858 PR>249 Pulse rate is greater than 249 bpm. (Nellcor
SpO
2
only)
8858 PR>239 Pulse rate is greater than 239 bpm.
(Masimo SpO
2
only)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals