Measurements (Accutorr Plus advanced models) Operation
3 - 22 0070-10-0692-02 Accutorr Plus™ Operating Instructions
2. Attach the appropriate corresponding Patient Cable (P/N 0012-00-1099-02, 0012-00-
1652, 0012-00-1599, or 0012-00-1653 from table above) to the sensor and plug the
other end of the patient cable into the SpO
2
connector (15).
NOTE: The PC Series Patient Cable is not used with the
LNOP
®
•DCSC Sensors.
NOTE: Do not place the sensor on an extremity with an invasive
catheter or blood pressure cuff in place.
NOTE: Ensure proper routing of patient cable to avoid
entanglement and/or strangulation.
CAUTION: When equipped with MASIMO® SpO2, use only MASIMO®
oxygen transducers including MASIMO LNOP® patient
dedicated adhesive sensors and MASIMO PC Series Patient
Cable. Use of other oxygen transducers may cause
improper Oximetry performance.
CAUTION: Tissue damage or inaccurate measurements may be caused
by incorrect sensor application or use, such as wrapping it
too tightly, applying supplemental tape, failing to inspect
the sensor site periodically, or failing to position it
appropriately. Carefully read the sensor directions for use,
the Accutorr Plus operating instructions, and all
precautionary information before use.
CAUTION: Excessive ambient light may cause inaccurate
measurements. Cover the sensor with opaque materials.
LNCS
®
•ADTX Adult single patient
adhesive sensor
0600-00-0121 > 30 kg. Disposable
LNCS
®
•PDTX Pediatric single patient
adhesive sensor
0600-00-0122 10 to 50 kg. Disposable
LNCS
®
•INF-L Infant single patient
adhesive sensor
0600-00-0123 3 to 20 kg. Disposable
LNCS
®
•NEO-L Neonatal single
patient adhesive sensor
0600-00-0124 < 3 kg. or
> 40 kg.
Disposable
LNCS
®
•NEO PT-L Neonatal preterm
single patient adhesive sensor
0600-00-0125 > 1 kg. Disposable
LNC-4 SpO2 Patient cable, 4' 0012-00-1652 All Re-usable
LNC-10 SpO2 Patient cable, 10' 0012-00-1599 All Re-usable
LNC-14 SpO2 Patient cable, 14' 0012-00-1653 All Re-usable
MASIMO
®
LNCS
®
Sensor Family
SELECTION PART NUMBER PATIENT SIZE
DISPOSABLE
/REUSABLE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray accutorr plus and is the answer not in the manual?

Mindray accutorr plus Specifications

General IconGeneral
BrandMindray
Modelaccutorr plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals