Accutorr Plus™ Operating Instructions 0070-10-0692-02 3 - 31
Operation How To Set The Clock (Date and Time)
4. Press the PATIENT INFO. Up or Down Arrow key (27 or 28) to change the
number.
Continue pressing the TIMER/TEMP key and the Arrow keys to set the month, day, and
year (in that order).
5. After the year has been selected, the next press of the TIMER/TEMP key (32) exits the
clock set mode and enters the new information.
To cancel a changed value while that value is still displayed, press the DEFAULTS key
(14) for less than 3 seconds.
If the TIMER/TEMP or Arrow keys have not been pressed for 15 seconds, the Accutorr
Plus returns to normal operation and saves any Time/Date changes.
When the clock is displayed, it displays real-time (current time). When the clock is displayed
while viewing previous data, frozen time is displayed. When frozen time is displayed, the
colon between the hours and minutes is illuminated continuously. When real-time is displayed
the colon between the hours and minutes flashes.