1 - 2
Do not place the equipment or accessories in any position that might cause it
to fall on the patient.
Do not start an infusion unless the setup was verified to be correct.
Clearing the occlusion result from line kinks, filter coagulation, etc. may
cause extra bolus to patients. Appropriate measures should be taken.
1.1.2 Cautions
CAUTION
When several infusion lines are connected to the same vascular access, there
may be back flow or prolonged response time of occlusion alarm. Therefore,
use check valve at the line end or follow local hospitals’ instructions while in
connection with other infusion system.
When using this equipment for enteral nutrition, do not use enteral fluids for
intravenous infusion to avoid patient injury, and use only dedicated
disposable enteral feeding sets for enteral nutrition.
Ensure that the equipment is supplied with continuous electric power during
work. Sudden power failure may cause data loss.
Electromagnetic fields may affect equipment performance. This makes it
necessary for other equipment used in the vicinity of this equipment to meet
EMC standards. Mobile phones, X ray and MRI equipment are all potential
interference sources because of their high-intensity electromagnetic
radiation.
Always install or carry the equipment properly to avoid damage caused by
drop, impact, strong vibration or other mechanical force. The equipment
should be observed to verify normal operation after fall, otherwise it cannot
be used.
Dry the equipment immediately in case of rain or water spray.
Some settings are password protected and can only be changed by
authorized personnel. Contact your department manager or biomedical
engineering department for the passwords used at your facility.
1.1.3 Notes
NOTE
The software was developed in compliance with IEC62304.
The equipment provides power-down storage. Alarms limit setting and
history record are saved and will be maintained if the equipment is powered
down suddenly. The storage time is equals to the equipment’s service life.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneFusion uVP and is the answer not in the manual?

Mindray BeneFusion uVP Specifications

General IconGeneral
BrandMindray
ModelBeneFusion uVP
CategoryWater Pump
LanguageEnglish

Related product manuals