3 - 3
• The positioning block of the drip chamber must be vertically inserted
through the positioning groove of the drop sensor.
• Do not excessively tilt the drop sensor, or expose it to direct sunlight during
infusion. Otherwise, accuracy of the drop sensor may be influenced.
• It is suggested that the signal line of drop sensor should be replaced every six
months.
3.4 Setting Up the Equipment
• Always use the accompanying power cord delivered with the pump.
• Before connecting the equipment to the AC mains, check that the voltage
and frequency ratings of the power line are the same as those indicated on
the equipment.
• Do not touch the power connector with wet hand. Eliminate the liquid or any
residue inside of or at the surroundings of the AC power input connector and
power cord connectors.
• Use the battery if the integrity of the protective earth conductor or the
protective earthing system in the installation is in doubt.
3.4.1 Charging the Battery
• The battery can only be charged by the pump or Dock.
• The battery is not charged when the pump is running at a rate higher than
1200ml/h.
• If the pump is run by battery power, ensure that the battery is adequately
charged.
• The pump refreshes the displayed date or time format of history record after
the date or time format is changed.