6.6 Occlusion Alarm
• If this pump is running at 0.1ml/h, and respectively configure the occlusion 
pressure alarm limit to the lowest level and highest level, the occlusion alarm 
delay time may reach up to one hour and 20.5 hours. Adjust the pressure 
limit to a lower level, or use the syringe pump for a low rate infusion.