6 - 2
6.2.1 Alarm Screen
NOTE
The alarm screen always displays the alarm of the highest priority.
6.3 Pausing Alarm Sound
NOTE
Except for the Battery Depleted alarm, the sounds of all alarms are paused by
pressing .
6.4 Nurse Call
NOTE
Do not rely exclusively on the nurse call system for alarm notification.
Remember that the most reliable alarm notification combines audible and
visual alarm indications with the patient’s clinical condition.
6.5 Alarm Solutions
WARNING
When an alarm occurs, check the pump’s status and handle the alarm as soon
as possible. If the alarms do not conform with the actual situation, contact
your service personnel.
NOTE
The pump stops infusion when a high priority alarm is triggered.
The pump continues infusion when a low priority alarm is triggered.
The pump stops infusion after the first Battery Depleted alarm occurs, and
the shutdown delay is at least three minutes.
Continue to work for at least thirty minutes after the first Low Battery alarm
occurs in the specified conditions (operating with a fully charged new battery
at 20°C ± 2°C, screen brightness configured to 2, default volume, Wi-Fi
disabled, drop sensor disconnected).