A.7 Infusion Specifications
Accuracy
Infusion accuracy: ≤ ±5%(use SHINVA ANDE single use 
infusion set for pump)
Infusion accuracy: ≤ ±4.5% (use B. Braun Intrafix 
Primeline
)
Bolus accuracy: ≤ ±5% or 0.02ml, whichever is greater 
Drip accuracy: ≤ ±10%
Note:Test in accordance with IEC60601-2-24:2012
Set range of the infusion rate/
purge rage/bolus rate 
Range of rate: 0.10ml/h to 2300ml/h (0.10 ml/h to 2000 
ml/h for blood transfusion)
Resolution:
0.01ml/h (0.10 to 99.99ml/h)
0.1ml/h (100.0 to 999.9ml/h)
1ml/h(1000 to 2300ml/h)
Occlusion pressure  15 levels selectable*: (50, 150, 225, 300, 375, 450, 525, 
600, 675, 750, 825, 900, 975, 1050, 1125)mmHg
Tolerance:
≤ ±75mmHg (for 50mmHg level, and operating at a rate
≤ 100 ml/h)
≤ ±20% or ±125mmHg, whichever is greater (for other 
pressure levels)
The maximum occlusion pressure is 1350mmHg. 
Bubble size 6 levels selectable: 15μl, 50μl, 100μl, 250μl, 500μl, 800μl
Minimum detectable bubble size: 15μl
Maximum volume (under 
single fault conditions)
≤ 0.5ml
KVO rate 0.1 to 5.0ml/h
Minimum resolution: 0.01ml/h
Time set range 00:00:01 to 99:59:59
VTBI set range 0.10 to 9999.99 ml
Resolution: 0.01ml
Weight set range 0.1 to 499.0 kg/0.2 to1100.1 lb
Drug Amt. set range 0.001 to 99999
Drug Amt. unit set range ng, μg, mg, g, mU, U, kU, EU, mmol, mol, mcal, cal, kcal,
mEq
Volume 0.10 to 9999.99ml
Conc. set range 0.001 to 9999.99