A - 6
• The infusion accuracy and pressure detection is affected by viscosity of
liquids and disposables used (for example diameter, material, elasticity and
needle).
• The infusion accuracy tests and occlusion pressure tests are performed in
accordance with IEC60601-2-24:2012 (test temperature: 20°C ± 2°C).
A.8 Recommended Infusion Sets
A.9 Occlusion Alarm Delay and Bolus Volume
Conc. unit set range ng/ml, μg/ml, mg/ml, g/ml, mU/ml, U/ml, kU/ml, EU/ml,
mmol/ml, mol/ml, mcal/ml, cal/ml, kcal/ml, mEq/ml
Dose Rate set range 0.001 to 99999
Product Name Type Manufacturer
Transfusion Sets for
Single Use
Transfusion SHANDONG WEIGAO GROUP MEDICAL
POLYMER CO., LTD
Single Use Infusion Set
for Pump
Regular SHINVA ANDE HEALTHCARE APPARATUS
CO., LTD
B.Braun Intrafix
Primeline
Precision B. Braun Melsungen AG
Rate (ml/h)
Occlusion alarm delay time (hh: mm: ss)
High occlusion alarm pressure
level
Low occlusion alarm pressure
level
1
< 01:17:26 < 00:04:37
25
< 00:02:43 < 00:00:14