A - 7
Test conditions:
■ Infusion set brand: B.Braun Intrafix Primeline
■ Anti-bolus: On
■ Test temperature: 20°C ±2°C
■
Infusion line length: 1 meter
• Occlusion alarm pressure, alarm delays and bolus volume may vary
depending on test conditions, temperature and tube length.
A.10 Infusion Accuracy Graphs
A.10.1 Infusion Accuracy at 1 ml/h
Rate
Bolus volume after occlusion (ml)
High occlusion alarm pressure
level
Low occlusion alarm pressure
level
25ml/h
< 0.18 < 0.18
Trumpet Curve at 1ml/h (2nd hour)
Observation window(min)
Percentage error of flow
Ep(Max)
Ep(Min)
Overall percentage error