• We recommend you to use an infusion set stated in this manual. If a non-
recommended infusion set must be used, perform the calibration and
performance test before use. Otherwise, the accuracy of the infusion and the
performance of the pump may be adversely affected.
• To ensure the accuracy of rate and alarm detection, the infusion set should be
calibrated in this pump before first use.
• When using the pump for blood transfusion, only use disposables dedicated
and labelled for transfusion.
• Take care that your hands are not squeezed when you close the pump door.
• This pump uses standard, single use infusion set with Luer lock connections.
• Single use accessories are not designed to be reused. Reuse may cause a risk
of contamination and affect the measurement accuracy.
• Make sure that the infusion set is located in both sides of the tubing channel
notches after loading the infusion set.
4.3 Starting Infusion
• Do not connect patient until disposables have been purged and loaded into
the pump. Connecting to patient before disposables are loaded and purged
can cause serious injury or death.
• Check that no drops are falling in the drip chamber before infusion starts or
stops. If drips are falling, close the roller clamp or the Robert clamp, do not
use the equipment, and contact your service personnel.
• If the infusion set type is Nutrition or the selected drug type is Enteral
Nutrition, “Air detection will be closed” is always displayed in the system
information area, “[Enteral feeding] Air detection will be closed” is displayed
before starting the infusion or bolus infusion.