3 Equipment Preparation
3.1 Equipment Preparation Safety Information
• Connect only approved devices to this equipment. Devices connected to the
equipment must meet the requirements of the applicable IEC standards (e.g.
IEC 60950 safety standards for information technology equipment and IEC
60601-1 safety standards for medical electrical equipment). The system
configuration must meet the requirements of the IEC 60601-1 medical
electrical systems standard. Any personnel who connect devices to the
equipment’s signal input/output port are responsible for providing evidence
that the safety certification of the devices has been performed in accordance
to the IEC 60601-1. If you have any questions, please contact Mindray
Scientific.
• If it is not evident from the equipment specifications whether a particular
combination with other devices is hazardous, for example, due to summation
of leakage currents, please consult the manufacturer or an expert in the field.
A determination must be made that the proposed combination will not
negatively affect the devices themselves or the patient's safety.
• Use only installation accessories specified by Mindray Scientific.
• The equipment should be installed by the authorized personnel.
• Ensure that the equipment is properly secured and positioned. Position
change and severe shock may lead to changes to the delivery accuracy.
• Before use, verify whether the packages are intact. In case of any damage, do
not apply it to patient.
• The equipment software copyright is solely owned by Mindray Scientific. No
organization or individual shall resort to modifying, copying, or exchanging
it or to any other infringement on it in any form or by any means without due
permission.