4 - 3
4.4 Purge
• If required, set the purge rate after the purge is started. The initial purge rate 
is 1200 ml/h.
• The Air in Line or Accumulated Air alarm will not given during purging.
• The volume used for purging is not added to the infused volume.
4.5 Bolus Infusions
• The delivered bolus volume will be added to the total infusion volume and 
subtracted from the volume to be infused (VTBI).
• If required, adjust the bolus rate in the BolusRate area during an automatic 
bolus infusion.
• The pump gives a beep every time a 0.5 ml bolus volume is infused.
4.6 Setting Keep Vein Open (KVO) Rate
• The pump runs for 30 minutes at a KVO rate. At the completion of the KVO 
infusion, the pump stops infusion, and gives a KVO Finish alarm.
• The volume used during KVO infusion will be added to the total infusion 
volume.
4.7 Replacing the IV Container
• Check the infusion status after replacing the IV container. If the pump is 
running a KVO infusion, reconfigure the infusion parameters before 
restarting the infusion.