4 - 4
4.8 Unloading the Infusion Set
WARNING
It is recommended to change the infusion set every twenty-four hours (or as
per national hygiene regulations or manufacturer’s instructions). If infusion
sets not recommended by this manual are used, adjust the fixing site every
four hours.
It is recommended to change the disposable enteral nutrition set every
twenty-four hours.
The accuracy of the infusion may be affected after long-term infusion.
To prevent free flow, ensure that the roller clamp or Robert clamp is closed
before removing the infusion set from the pump.
4.9 Viewing the Infused Volume
NOTE
The infusion volume cannot be cleared when an infusion is running.
4.10 Turning Off the Pump
CAUTION
Press and hold the power switch for no less than 10 seconds to forcibly shut
down the pump if it could not be shut down normally. This may cause loss of
patient data.
NOTE
Turning off the pump does not disconnect the pump from the AC mains. To
completely disconnect the power supply, unplug the power cord.