4 - 1
4 Getting Started
4.1 Setting Up the Pump
• Check that visual and auditory alarm signals are presented correctly when 
the equipment is powered on. Do not use the equipment if you suspect it is 
not working properly, or if it is mechanically damaged. Contact your service 
personnel or us.
• Before putting the system into operation, the operator must verify that the 
equipment, connecting cables and accessories are in correct working order 
and operating condition.
• Stay within 1 meter (39 inches) of the pump while setting it up and operating 
it, making sure that you have a clear view of the pump interface.
• The equipment uses a mains plug as isolation means to the mains power. Do 
not locate the equipment in a place difficult to operate the mains plug.
4.2 Loading the Infusion Set
• To ensure the accuracy of air bubble detection, check and remove the 
remained fluid in the infusion set slot before loading the infusion set.
• While loading the infusion set, do not touch the anti free-flow clamp to avoid 
being hurt.
• The pump must be mounted to the same level as the patient’s heart. The most 
accurate pressure monitoring in the infusion set is achieved when the pump 
is positioned close to the patients heart level.