A - 1
A Product Specifications
A.1 Classifications
The equipment is classified, according to IEC60601-1:
A.2 Environmental Specifications
Storage Conditions: Corrosive-free and ventilated
• The pump may not meet the performance specifications if stored or used 
outside the specified temperature and humidity ranges. If the performance 
of the equipment is degraded due to aging or environmental conditions, 
contact your service personnel.
Type of protection against 
electrical shock
CLASS I EQUIPMENT, equipment energized from an 
internal electrical power source.
Degree of protection against 
electrical shock
Defibrillation-proof type CF applied part (direct cardiac 
application)
Mode of operation Continuous
Degree of protection against 
harmful ingress of water
IP33
Degree of safety of application 
in the presence of flammable 
anesthetic mixture with air or 
with oxygen or nitrous oxide
The equipment is not suitable for use in the presence 
of a flammable anesthetic mixture with air or with 
oxygen or nitrous oxide
Degree of mobility Portable
Item
Temperature 
(ºC)
Relative humidity 
(noncondensing)
Barometric (kPa)
Operating conditions 5 to 40 15% to 95% 57.0 to 107.4
Storage conditions –30 to 70  10% to 95% 16.0 to 107.4