5 - 1
5 Using the Dock
5.1 Securing the Pump in the Dock
• The alarm sound from the pump is disabled when the pump is secured in the 
BeneFusion nDS Infusion Supervision System (Dock). The alarm sound is 
given by the Dock. 
5.2 Relay Infusion
• Relay infusion is available for Rate Mode, Micro-infusion Mode, Time Mode, 
and Dose Mode.
• When a relay infusion is set up, the sequence of the current pump in the rely 
is displayed in the system information area. For example, symbol   
indicates that the current pump is the second in a circular relay.
• You cannot change the sequence of the pumps when the relay infusion is set 
up.
 
• Removing a relay pump from the Dock cancels the relay infusion.
• For a circular relay, initiating a middle pump cancels the relay. For a 
customized relay, initiating a middle pump removes the pumps before it 
from the relay.