8 - 2
• If you do not configure the loading dose parameters, the pump infuses at the
Primary Rate until the set VTBI is finished.
8.5 Micro-infusion Mode
• Rate setting is necessary for an infusion, while time and VTBI are optional in
the micro-infusion mode.
8.6 Intermittent Mode
• Total VTBI and Maintain Rate are optional parameters. If the Maintain Rate is
not set, infusion stops at the maintenance stage. If the Total VTBI is not set,
the infusion stops when the IV container is empty.
8.7 Ramp Mode
• The Steady Rate can only be obtained by calculation. It is not available for
manual input.
• Up Time and Down Time are optional parameters. The pump runs an infusion
at the steady rate if they are not set.
8.8 Dose Time Mode
• In the dose time mode, the supported dose rate units are X/min, X/h, and X/
24h, in which X represents ng, ug, mg, g, mU, U, kU, EU, mmol, mol, mcal, cal,
kcal, and mEq.
• Rate and time can only be obtained by calculation. They are not available for
manual input.