21 - 6 Defibrillator/Monitor Operator’s Manual
Lead Set 3-Lead, 5-Lead 3-Lead Sets the ECG lead type.
This setting affects the default
waveform position of the
[Waveform Setup] menu.
QRS Volume 0 to 10 2 This setting is linked with the
[QRS Volume] setting in the
[SpO2 Setup] menu
ECG1 3-Lead: I, II, III II Available options are defined by
the current setting of [Lead
Set].
5-Lead: I, II, III, aVL, aVR, aVF,
V
ECG2 3-Lead: I, II, III / • If 3-lead is used, the default
setting of this item is blank
and this item is inactivate.
• The options of [ECG1] are
not available for [ECG2].
5-Lead: I, II, III, aVL, aVR, aVF,
V
V
Sweep 6.25 mm/s, 12.5 mm/s, 25
mm/s, 50 mm/s
25 mm/s Sets the ECG waveform speed.
HR Alarm On, Off On /
HR Alm Lev High, Med Med /
HR High Adu (Low+2) to 300 120 Step: 1 bpm
Ped (Low+2) to 300 160
Neo 200
HR Low Adu 15 to (High-2) 50
Ped 75
Neo 100
Arrhythmia On, Off Off Sets whether to enable the
arrhythmia analysis.
ARR Alarm On, Off On /
Menu Item Options/Range Default Description

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)