Defibrillator/Monitor Operator’s Manual 21 - 7
ARR Alm Lev PVCs/min High, Med, Low Med /
R ON T
VT>2
Couplet
PVC
Vent Rhythm
Bigeminy
Trigeminy
Tachy
Brady
PNP
PNC
Missed Beat
A-Fib
Multif. PVCs
Nonsus. Vtac
Pause
Irr. Rhythm
Asystole Delay 3 to 10 5 /
V-Tach Rate 100 to 200 130 /
V-Tach PVCs 3 to 99 6 /
Vbrd Rate 15 to 60 40 /
Vbrd PVCs 3 to 99 5 /
Extreme Tachy Adu 60 to 300 160 /
Ped 60 to 300 180 /
Extreme Brady Adu 15 to 120 35 /
Ped 15 to 120 50 /
PVCs High 1 to 10 10 /
Tachy Adu 60 to 300 120 /
Ped 60 to 300 160 /
Brady Adu 15 to 120 50 /
Ped 15 to 120 75 /
Multif. PVCs Window 3 to 31 15 /
Pause Time 1.5, 2.0, 2.5 2.0 /
Menu Item Options/Range Default Description

Table of Contents