E - 4 Defibrillator/Monitor Operator’s Manual
CO2 CO2 Sensor High Temp L C Check, stop using or replace the sensor.
CO2 No Watertrap L B Check the watertrap connections.
CO2 Zero Failed L A Check the CO
2
connections. After the sensor’s
temperature becomes stabilized, perform a zero
calibration again.
CO2 Module Error L C
There is a problem with the CO
2
module
, or a problem
with the communications between the main unit and
the CO
2
module. Restart the equipment.
CO2 Occlusion L C The airway was occluded. Check the airway and remove
the occlusion.
CPR sensor CPR Sensor Err H C There is a self-test error or communication problem with
the CPR sensor. Contact your service personnel.
CPR Sensor Low Battery M C The battery power of the CPR sensor is low. Charge the
battery by connect the CPR sensor to the equipment.
CPR Sensor Need
Service
H C The compressions using the CPR sensor exceed the
expected numbers. Contact your service personnel.
CPR Sensor Cable Fault L C An error occurred to the CPR sensor cable. Replace the
CPR sensor cable.
Change CPR Sensor
Battery
L C The CPR sensor battery is aging. Contact your service
personnel.
CPR Sensor Bat. Charge
Err
L C The CPR sensor cannot be charged. Contact your service
personnel.
Main control
system
No Speaker L C Make sure that the speaker is connected.
Power Board Comm Err H C
An error occurred to the power board, or there is a
problem with the co
mmunications between the power
board and the main unit. Restart the equipment.
Keyboard Comm Err L C
An error occurred to the
keypad board, or there is a
problem with the communications between the keypad
board and the
main unit. Restart the equipment.
Therapy Module Comm
Err
S C
An error occu
rred to the therapy module, or there is a
problem with the communications between the therapy
module and the
main unit. Restart the equipment. If the
proble
m persists, contact your service personnel.
Main Control Selftest
Err
H C The main control voltage is abnormal. Replace the main
control board.
Wifi Module Fault L C Contact your service personnel.
Machine Type Error H C
RT Clock Need Reset L C Reset system time.
RT Clock Err H C An error occurred to the RTC chip, or the button cell is
depleted. Replace corresponding part.
Memory Error L C There is a problem with the data card. Format the CF card.
If the problem persists, contact your service personnel.
Last User Test Failed L C Run a successful user test.
Last Auto Test Failed L C Run a successful user test again.
NO CMS. L C The equipment is disconnected from the CMS. Check the
network connection.
IP Address Conflict L C Network IP conflicts. Check the network settings.
Measurement Alarm Message L I Cause and solution

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
TypeDefibrillator
DisplayColor TFT LCD
Display Size5.7 inches
WaveformBiphasic Truncated Exponential (BTE)
Heart Rate Range15-300 bpm
SpO2 Range0-100%
NIBP Measurement MethodOscillometric
Printer TypeThermal
Paper Width50 mm
Paper Speed25 mm/s
IP RatingIP44
ModesManual, AED, Pacing
AED Mode Energy Sequence150J, 200J
Charging Time<5 seconds to 200J
Battery TypeLithium-ion
ECG Monitoring3-lead
Energy Selection1-360 J (Manual Mode)