Structure and Assembly/Disassembly 9-47
1 M10X25 inner head screw and washer
3. Put the rear cover of monitor near to the installation board of monitor, then move it downwards
to insert the installation board into two hanging holes and two hanging bands, then use 6
M4X12 cross panhead screws to fix the monitor.
4. Visually check lifted/ lowered functions: make sure the monitor, upper arm/ lower arm can
move normally when monitor is turned lifted/ lowered (-90,+20) degree, upper arm (-45,
+10)degree, lower arm(-45,+25)degree. No abnormal sound exists and the monitor can
keep still in each position.
5. Check rotation function: each part can be rotated left/ right normally, the rotating angle of
monitor and lower arm are about -90~+90 degree. No abnormal sound exists and each part
can be moved smoothly.
6. Adjust the lower arm in -45 degree and upper arm in +10 degree direction, then guide two
cables of monitor into the connecting base from rotating axis of lower arm, and come out from
rotating axis to the parallel cable groove of upper arm(Guide the signal cable first, then
power line, tidy up two cables, too).
Come out from rotating axis of upper arm
Guide two cables into the connecting base from rotating axis of lower arm

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals