Structure and Assembly/Disassembly 9-49
10. Adjust the lower arm and turn anticlockwise about 90 degree, adjust upper arm in -45 degree
direction, lower arm in +25 degree direction, the supporting arm is the position in which
interior cables are the longest at this time.
Upper arm in -45 degree Lower arm in +25 degree
Turn anticlockwise about 90 degree
11. Two cables are in the cable groove of upper arm: cables in top block of upper arm are
loosened and not too long seeing from lateral. Cables are close to the interior wall of
connecting base.
12. Arrange two cables of monitor into the parallel cable groove of lower arm, then use two cable
fixing blocks and 2 M3X8 cross panhead combination screw to fix the cables in the groove.