5-10 Image Optimization
Dual Live
Description
Display different image effects of one probe for a better observation.
Operation
Touch [Dual Live] on the touch screen to turn on/off the function, and dual-split
window of images are displayed on the main screen.
Two pages of adjustable parameters are displayed in the touch screen as well;
where, shared parameters and left window parameters are displayed in the B(L)
page, while right window parameters are displayed in the B(R) page.
In the image parameter area on the upper right corner of the main screen,
parameters of the both windows are displayed.
Impacts
Image magnification is available in dual live mode.
LGC
Description
Adjust gain of scan lines to increase the image lateral resolution.
Operation
Touch [LGC] on the touch screen to enter the adjusting dialogue box.
The 8 rods displayed on the touch screen indicate the corresponding image areas
on the main screen.
Touch the to adjust the gain, move downwards to decrease the gain, while
move upwards to increase the gain.
In addition, the system provides several preset parameters for imaging.
Echo Boost
Description
This function can improve contrast and decrease noises, so that a much clearer
boundary can be seen.
Operation
Touch [Echo Boost] on the touch screen to turn on the function (when the function is
turned on, the [Echo Boost] button is highlighted in green and the system has Echo
Boost status indicating on the image parameter area).
Impacts
This function is available for using phased probes in cardiac exam mode and using
linear probes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 and is the answer not in the manual?

Mindray DC-80 Specifications

General IconGeneral
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Tissue Doppler, 3D/4D Imaging
ProbesConvex, Linear, Phased Array, Volume
Advanced FeaturesAuto Optimization, Auto EF
ConnectivityDICOM 3.0, USB, HDMI, VGA, Ethernet
Transducer Ports4
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Small Parts, Musculoskeletal, Urology, Pediatrics, Neonatal
Transducer TypesConvex, Linear, Phased Array, Volume
Power Supply100-240V AC, 50/60Hz
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity30% to 80%

Related product manuals