5-14 Image Optimization 
5.4  Color Mode Image Optimization 
The Color mode is used to detect color flow information, and the color is designed to judge the 
direction and speed of blood flow. 
Generally, the color above the color bar indicates the flow towards the probe, while the color below 
the color bar indicates the flow away from the probe; the brighter the color, the faster the flow 
speed, while the darker the color, the slower the flow speed. 
5.4.1  Basic Procedures for Color Mode Imaging 
1.  Select a high-quality image during B mode scanning, and adjust to place the area of interest in 
the center of the B mode image. 
2.  Press <Color> to enter B + Color mode. Roll the trackball to change position and size of the 
Region of Interest (ROI) and press the <Set> key to set. 
3.  Adjust the image parameters during scanning to obtain optimized images. 
4.  Perform other operations (e.g. measurement and calculation) if necessary. 
5.4.2  Color Mode Image Parameters 
  In Color mode scanning, the image parameter area in the upper right corner of the screen 
displays the real-time parameter values as follows: 
 
Pulse Repetition Frequency (PRF) 
 
  During Color mode imaging, menus of image optimizing for B mode and Color mode are 
displayed on the touch screen at the same time. You can switch between the 2 modes by 
clicking the mode tabs. 
  In Color mode, acoustic power is synchronous with that of B mode. Adjustment of the depth to 
the B mode image will lead to corresponding changes in Color mode image. 
5.4.3  Color Mode Image Optimization 
Color Gain 
Refers to the overall sensitivity to flow signals, and this function is used to adjust 
the gain in Color mode. The real-time gain value is displayed in the image 
parameter area in the upper right corner of the screen. 
Rotate the <Color> knob clockwise to increase the gain, and anticlockwise to 
decrease. 
The adjusting range is 0-100. 
Increasing the gain will increase the flow signal presented as well as noise, while 
the signals may be missing when the gain is adjusted too low.