5-22 Image Optimization
T/F Res
This function is used to adjust for a balance between time resolution and
spatial resolution.
Adjust through the [T/F Res] item on the image menu.
There are 0-4 levels of T/F Res values available.
Dynamic Range
The dynamic range conveys the information that being transformed from echo
intensity to gray scale.
Adjust through the [Dynamic Range] item on the image menu.
The adjusting range is 24-72dB in increments of 2dB.
The more the dynamic range, the more specific the information, and the lower
the contrast with more noise.
HPRF
HPRF mode is used when detected velocities exceed the processing
capabilities of the currently selected PW Doppler scale, or when the selected
anatomical site is too deep for the selected PW Doppler scale.
Turn on or off the function through the [HPRF] item on the image menu.
HPRF enhances the range of detecting high-velocity flow.
Speed
This function is used to set the scanning speed of PW mode imaging.
Click the [Speed] item on the image menu.
There are 1-6 levels of scan speed available, the smaller the value the faster
the speed.
Changing the speed makes it easier to identify the cardiac cycles and to
detect more details.
Display Format
To set the display format of PW mode image with B mode image.
Click the [Display format] item on the image menu to adjust.
Options: V2:1, V1:2, V1:1, Full.
Audio
This function is used to adjust the output audio in spectrum map.
Click the <Audio> item on the image menu to adjust the volume.
The adjusting range of the audio is 0-100% in increments of 2.
Utilizing the output audio helps to identify the feature and status of flow.
Gray Map
This function applies the gray correction to obtain the optimum images.