6. Operating | 6.3 - Other functions via keypad | 143
7. Press "OK".
Pressing the key increases the counting in line with the parameter
settings. Every time the "OK" key is pressed, 1,000mA are added to
the output (up to a maximum of 20,000mA).
8. Press "OK" again to repeat steps 7 and 8.
The analog output is faulty if the analog output current cannot be
measured as described.
9. Press "Exit" to exit the test menu.