6. Operating | 6.2 - Configuration and adjustment via the keypad | 106
6.2.5 Displaying adjustment data of the internal WP
WP = weighing point
Procedure
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Press "OK" to select the menu item.
Calib appears on the display. When the CAL switch is closed, CAL.CLS
appears on the display (continue with step 7).
4. Press "OK".
nEW appears on the display.
5. Press until VIEW appears.
6. Press "OK".
CAL.oPn (CAL switch open) appears on the display.
7. Press "OK".
MAX.FSd (max. capacity of the weight display = full scale deflection)
appears on the display.
8. Press "OK".
The set max. capacity appears on the display.
9. Press "OK".
Cd 002 appears on the display.
10. Press "OK".
StEP appears on the display.
11. Press "OK".
The set scale interval appears on the display.
12. Press "OK".
Cd 003 appears on the display.
13. Press "OK".