6. Operating | 6.2 - Configuration and adjustment via the keypad | 99
6.2.3 New adjustment of the internal weighing point
Example:
Max (max. capacity) 600 kg
Scale interval 0.2 kg
Set dead load with empty scale.
Note:
When a new adjustment is started, the device has the following default
settings:
Max 3000 kg
Scale interval 1 kg
Start
1. Switch on the device, see ChapterSwitching on the device. Observe the
warm-up time of the device!
2. Open the CAL switch, see ChapterCAL switch.
3. Press and hold for 2–3 seconds.
SEtuP appears on the display.
New adjustment (Cd 000)
4. Press "OK".
Cd 000 appears on the display (Cd flashes).
5. Press "OK".
Calib appears on the display.
6. Press "OK".
nEW (for new adjustment) appears on the display.
Defining Max (max. capacity) and weight unit (Cd 001)
7. Press "OK".
Cd 001 appears on the display.
8. Press "OK".
MAX.FSd (max. capacity of the weight display = full scale deflection)
appears on the display.
9. Press "OK".