6. Operating | 6.2 - Configuration and adjustment via the keypad | 100
------.kg appears on the display.
10. Press to move the decimal point one place to the left -----.-kg.
11. Press "OK".
00300.0kg appears on the display (3 flashes).
12. Press (3×) to set the number 6 for 600.0 kg.
13. Press (4×) to select the weight unit (kg flashes) and change using if
necessary.
Defining the scale interval (Cd 002)
14. Press "OK".
Cd 002 appears on the display.
15. Press "OK".
StEP appears on the display.
16. Press "OK".
1 appears on the display.
17. Press to change the verification interval to 2 for 0.2 kg.
Defining the dead load (Cd 003)
18. Press "OK".
Cd 003 appears on the display.
19. Press "OK".
dEAdLo appears on the display.
20. Press "OK".
LoAd (dead load with empty scale) appears on the display.
21. If necessary, press to select MVoLt (mV/V input).
Continue for "dead load with empty scale":
22. Press "OK".
unLoAd appears on the display.
23. Clear the scale.
24. Press "OK".