4. Device installation | 4.3 - Hardware construction | 78
4.3.10 EtherNet/IP interface
The EtherNet/IP fieldbus is a complete EtherNet/IP adapter (slave) for
network connection.
The interface can be configured by front keys.
Technical data
Description Data
Internal connection
Pin strip,
50‑pin
Transfer rate 100Mbit/s, full duplex
Protokoll EtherNet/IP
Connection mode Network
Configuration EDS file "sag_5310_EthernetIP-2P.eds"
Potential isolation yes
Cable type Twisted pairs, screened, e.g., patch cable CAT5 Au
tolink (straight or crossover)
Cable impedance 150Ω
Cable length to HUB max. 115m
Certificate
EtherNet IP Specification ODVA File No. 10286
Test Date: 06.09.2005
Vendor ID 90
Note:
The EDS file is available to download online:
http://www.minebea-intec.com