4. Device installation | 4.3 - Hardware construction | 79
4.3.10.1 LEDs
Identif. Description
NS Network status LED
MS Module status LED
grn Green: flashes when
there is data traffic (ac
tivity)
yel Yellow: lights up when
there is an existing con
nection (link)
Network status (NS)
LED status Description Comments
Off Module is offline
No power
No IP address
Constant green Module is online One or more connections exist (CIP
class1 or3).
1Hz flashing green Module is online No connection
Constant red Critical connection error Duplicate IP address
1Hz flashing red Connection time limit For one or more connections (CIP
class1 or3)